kic2023炉温测试仪从炉子里出来读不出数据怎么办?
<p>问题:kic2023炉温测试仪从炉子里出来读不出数据怎么办?<p>答案:↓↓↓<p class="nav-title mt10" style="border-top:1px solid #ccc;padding-top: 10px;">康欣的回答:<div class="content-b">网友采纳 通常是仪器过热自动保护后怎样进行复位? 退出KIC2000软件 1,找到KICProfiler2000Logkic2000usersettings.kiccfg,用记事本方式打开 2,把文件中的ViewHostDialog=0改为=1,然后保存并退出. 再次启动KIC2000软件你就可以看到KICHost界面了. Cantrythesesteps,remarktousingEnglishLanguage Recommendisuse9VALKALINEBattery 1.x09MakesuretheSlinKIC2000softwareisclose. 2.x09Undertherootdirectoryonyourharddrive,underthefoldertheKICProfiler2000youwillfindafoldercalled"Log".(UsingWindowsExplorer,opentheX:KICProfiler2000Logkic2000usersettings.kiccfgfile) 3.x09UndertheLogfolderthereisafilecalled"KIC2000UserSettings.KICCFG" 4.x09OpenthisfileusingNoteorWordPadtexteditor. 5.x09Withfileopenedunderthetexteditor,findtheline:("ViewHostDialog=0",Setthislineto=1,Savechangestothefile.) 6.x09RestarttheSlimKIC2000softwareandyoumoticethatasecondapplicationwillopen,theKICHost.(OntheWindowsTaskbar,clickonthe"KICHost"tab) 7.x09DirectconnectiontoSlimKIC2000,restartSlimKIC2000Unit. 8.x09Select"ConfigureSlimKIC". 9.x09AnotherWindowwillopen.Select"WriteSlimKICSetuop".Youshouldseedialogappearinthatwindowwhichsays"WriteSlimKICSuccessful". 由深圳市时汉电子为你专业服务,如有不明白的地方.CALL:13802708085唐/R